不同根尖定位仪和测量锉直径对测量根管工作长度准确性的影响

Research on the Accuracy of Measuring Canal Working Length by Different Diameter of Files and Electronic Apex Locators

【作者】 李勋

【导师】 张志民;

【作者基本信息】 吉林大学 , 口腔医学, 2010, 硕士

【摘要】 根管工作长度测量的准确性是决定根管治疗成功的一个重要因素,是根管预备、根管充填的基础和前提。目前获得根管工作长度的的方法主要有X线比例计算法和根尖定位仪法等。其中根尖定位仪法是一种通过测定根尖孔牙周膜与口腔粘膜的电阻值来确定牙根管长度的方法,应用的仪器为根尖定位仪,因其准确度高、操作简单、减少患者的X线的照射次数等优点被广泛应用。根尖定位仪经过半个多世纪的发展,现在临床上常用的第三代(双频交流信号)和第四代(多频交流信号)产品与第一代相比准确性也有较大提高。但不同原理的根尖定位仪和根尖孔的大小之间关系仍是影响测量准确性的关键因素。为此,本实验采用磷酸盐缓冲液琼脂凝胶构建模拟牙齿在体内环境的模型,分别应用Root ZX100、Raypex5、ProPex和APEX-EYE KM-9000根尖定位仪测量根管工作长度,利用Digora系统直接数字化显影得到离体牙工作长度的真实值作对比,发现不同原理的根尖定位仪测量根管工作长度的准确性与测量锉直径有密切的相关性,所得结果为临床上更准确使用根尖定位仪法测量根管工作长度提供实验依据。

【Abstract】 Root canal therapy(RCT) is an efficient method of treating pulposis and periapical disease. Root canal preparation is the most important step in RCT and it is the precondition and foundation of root canal disinfection and filling.Working length mentioned in RCT means the distance from the reference point on the crown to the apical stop and accurate working length is helpful to remove infected mass inside root canals effectively and protect apical constriction from being ruined by canal preparing instruments which are beyond root foramens.Accurate working length plays an important role in canal cleaning and shaping which induce successful endodontic treatments.At present,there are many methods to determine the working length such as tactile perception,radiographic method and electronic apex locator method.As X-ray film shows two-dimensional images of objects and the dentino-cemental junction is determined by the apical plane in radiographic method,it is hard to determine apical constriction with a vague X-ray film.In recent years digital X-ray imaging technique makes root images more distinct and after clinical trials, better accuracy of working length is obtained by radiographic method with digital X-ray imaging technique.Nowadays electronic apex locator method is most commonly used to determin the working length in the root canal therapy. After half a century’s development, the accuracy obtained by electronic apex locators of the third and fourth generation is better than that obtained by electronic apex locators of the first generation.The advantages of electronic apex locator are as bellow: accurate, simple, convenient and less X-ray radiation to patients. Although the electronic apex locator presents the working length with high accuracy, the result would be affected by several factors. The hot spot of current studies is about the influence of the root canal apical foramen diameter and the solution quality inside the canal on the measuring accuracy and it is concluded that the residual pulp and strong electrolyte solution inside root canals can affect the measuring results and the measuring accuracy decreases when the apical foramen is larger than 0.4㎜ in many reaserches. Electronic apex locators with different working principles and apical foramens of different diameters are key factors in testing working length.The objective of this study is to discuss the selection of files when measuring the working length of root canals prepared regularly and compare the accuracy of 4 electronic apex locator species commonly used in clinic.40 isolated teech are selected as subjects of this study after being screened.Each of the teeth has only one straight cannal.After the pretreatment of the teeth, model was established with teeth and phosphate buffer agar gel,and then put initial file into cannal to conduct digital imaging and gain working length.In order to make sure the value,we conduct digital imaging again,until the tip of the file locate at the apical narrow region,the length of the file in cannal equals to the standard value of working length.Then enlarge the cannal with Mtwo nickel-titanium file from 10# untile 30# and measure the working length with 8#-30# K files and 4 kinds of electronic apex locator.Analyze the results statistically to evaluate the differences between the measurement and the standard value.Finally results are found as follows:1.Group Root ZX:The mean value between the standard and 8#,10#,15# K file measurements differs significantly whileno statistical differen-ce is found between the mean value of the standard and20#,25#,30#K file measurements. Group Raypex:The mean value between the standard and 8#,10# K file measurements diffe-rs significantly while no statistical difference is found between the meanvalue of the sta-ndard and 15#,20#,25#,30# Kfile measurements.Group ProPex:The mean value between the standard and 8# K file measurements differs significantly while no statistical difference is found between the meanvalue of the standard and 10#,15#,20#,25#,30# K file measurements.Group APEX-EYE: The mean value between the standard and 8# K file measurements dif-fers significantly while no statistical difference is found between the meanvalue of the standard and 10#,15#,20#,25#,30# K file measurements. 2.The accuracy of Root ZX is 85%; the accuracy of Propex is 87.5%;the accuracy of Reypex is 90%;the accuracy of APEX-EYE is 90%.It is found that all the files of different diameters affect the results measured by the four electronic apex locator species.The results obtained by files of significantly different diameters from those of root canals are statitically smaller while the results obtained by files of diameters similar to those of root canals are of no statistical significance.Evaluated by the standard value(from the apical constriction±0.5mm to the reference point on the crown),the accuracy sequence of measuring results is Root ZX<Propex<Raypex=APEX-EYE,which is coincident with researches done.

  • 【网络出版投稿人】 吉林大学
  • 【网络出版年期】2010年 09期
  • 【分类号】R781.05
  • 【被引频次】1
  • 【下载频次】171
节点文献中: 

本文链接的文献网络图示:

本文的引文网络