全自动红外测油仪的研制

Development of fully automatic infrared oil measuring instrument

  • 推荐 CAJ下载
  • PDF下载
  • 不支持迅雷等下载工具,请取消加速工具后下载。

【作者】 郭波张森林振强刘传荣郑鹏胡德栋隋峰许爱华王新军

【Author】 GUO Bo;ZHANG Sen;LIN Zhenqiang;LIU Chuanrong;ZHENG Peng;HU Dedong;SUI Feng;XU Aihua;WANG Xinjun;Shandong Institute of Metrology;Taian Kerui Optical Instrument Co., Ltd.;Qingdao University of Science and Technology;

【通讯作者】 郭波;

【机构】 山东省计量科学研究院泰安市科瑞光学仪器有限公司青岛科技大学

【摘要】 研制新型全自动红外测油仪。通过对进样系统、萃取系统、吸附过滤系统和检测系统自动化和集成化设计,实现了对水中油分的全自动检测。仪器的测量范围为0~200 mg/L。自动进样器取样体积误差在1%以内,仪器示值误差在样品浓度不大于10 mg/L时为0.1 mg/L,样品浓度大于10 mg/L时为0.9%,仪器重复性为0.2%(n=6),零点漂移为–0.1 mg/L,示值漂移为0.6%,最小检出浓度为0.1 mg/L,两种水样测量结果的相对标准偏差分别为1.28%和0.91%(n=11),加标回收率分别为96.8%和96.1%,检出限为0.006 mg/L。该仪器测量准确,性能稳定,能够降低工作强度,提高工作效率,减少有机物危害,为环境监测提供有效保障。

【Abstract】 A new fully automatic infrared oil detector was developed. By the automatic and integrated design of the injection system, extraction system, adsorption filtration system and detection system, the automatic detection of oil in water was realized. The measuring range of the instrument was 0–200 mg/L. The autosampler volume error was less than 1%. The indication error of the instrument was 0.1 mg/L when the concentration of sample was less than or equal to 10 mg/L, and the indication error was 0.9% when the concentration of sample was more than to 10 mg/L. The instrument repeatability was 0.2%(n=6). The zero drift was –0.1 mg/L. The indicated value drift was 0.6%. The minimum detection concentration was 0.1 mg/L. The relative standard deviations of the measured water samples were 1.28% and 0.91%(n=11). The spiked recoveries were 96.8% and 96.1%. The detection limit was 0.006 mg/L. The instrument has accurate measurement and stable performance. It could reduce work intensity, improve working efficiency, reduce the harm of organic hazards, and provide effective guarantee for environmental monitoring.

  • 【文献出处】 化学分析计量 ,Chemical Analysis and Meterage , 编辑部邮箱 ,2019年03期